Publication Type Journal Article
Title Adaptation of a commercial ion selective fluoride electrode to a tubular configuration for analysis by flow methodologies
Authors António C. L. Conceição Margarida Correia dos Santos M. L. S. Simoes Goncalves
Groups
Journal TALANTA
Year 2008
Month June
Volume 76
Number 1
Pages 107-110
Abstract The determination of fluoride ions in water samples is accomplished by using a tubular flow through detector constructed by drilling a channel through a commercially available LaF3 crystal electrode in such a way that the original contacts of the non-modified unit are maintained. Its performance when incorporated in both FIA and SIA systems was evaluated and the results show that the tubular unit retains the characteristics of the non-modified electrode. In SIA conditions an extended linear range of response and lower detection limit were achieved when compared with the electrode performance in FIA conditions. These aspects together with the additional advantage of low sample and reagent consumptions in SIA when compared to FIA, makes the incorporation of the proposed tubular ISE in a SIA system the preferred approach for on line determination and monitoring of fluoride content in natural water samples. (C) 2008 Elsevier B.V. All rights reserved.
DOI http://dx.doi.org/10.1016/j.talanta.2008.02.010
ISBN
Publisher ELSEVIER SCIENCE BV
Book Title
ISSN 0039-9140
EISSN
Conference Name
Bibtex ID ISI:000256934200019
Observations
Back to Publications List