Publication Type Journal Article
Title Novel Solid-Contact Ion-Selective Microelectrodes for Localized Potentiometric Measurements
Authors Svetlana V. Lamaka Maryna Taryba M. L. Zheludkevich M. G. S. Ferreira
Groups
Journal ELECTROANALYSIS
Year 2009
Month November
Volume 21
Number 22
Pages 2447-2453
Abstract The design and properties of novel type of solid-contact ionophore-based ion-selective microelectrodes are reported. The microelectrode is based on an insulated needle-shaped metallic wire with an exposed apex. The ion-to-electron transducer is made of poly(3-octylthiophene-2,5-diyl) and placed between an ion-selective membrane and the metallic tip. The ion-selective polyvinyl chloride-based membrane is deposited atop the layer of conductive polymer. The length of the ion-sensitive part of the electrode is less than 10 mu m. pH and Mg2+-selective microelectrodes were constructed and tested showing stable potential and fast response that are essential properties for the practical application of microelectrodes for localized scanning measurements.
DOI http://dx.doi.org/10.1002/elan.200900258
ISBN
Publisher
Book Title
ISSN 1040-0397
EISSN
Conference Name
Bibtex ID ISI:000272390100007
Observations
Back to Publications List