Title Accuracy and precision of heat capacity measurements using a heat flux differential scanning calorimeter
Authors TMVR deBarros RC Santos Anabela Fernandes M. E. M. Piedade
Groups MET
Journal THERMOCHIMICA ACTA
Year 1995
Month
Notice: Undefined index: in /afs/ist.utl.pt/groups/cqe/web/tmp/templates_c/77f86a5f762542dadf50c7f7fefa96acd45c2726_0.file.paper.tpl.html.php on line 163
Volume 269
Number
Pages 51-60
Abstract The accuracy and precision of a Setaram DSC 121 for the measurement of heat capacities of solids in the continuous mode was tested using sapphire and benzoic acid. The influence of the heating rate, mass of sample, and position of the crucible in the measuring tube on the obtained results is discussed. In the temperature range studied (320-440 K for sapphire and 320-370 K for benzoic acid) the accuracy and precision of the determination, based on the average results of three independent experiments, are better than 1\% provided that adequate experimental conditions (large sample masses and low heating rates) and data handling procedures are used.
DOI http://dx.doi.org/10.1016/0040-6031(95)02706-8
ISBN
Publisher
Book Title
ISSN 0040-6031
EISSN
Conference Name
Bibtex ID ISI:A1995TQ42700006
Observations